The RKS-01V-SOLO radiometer-dosimeter is designed for radiation monitoring of the environment, workplaces and vehicles during inspections carried out by teams of the Ministry of Defense, the Ministry of Emergencies, the Ministry of Internal Affairs, intelligence services, customs and border control authorities, and the army.
The RKS-01V-SOLO can be programmed with preset ADER threshold levels of gamma X-ray radiation, neutron radiation, as well as surface flux density of beta and alpha particles. The device has automatic selection of ionizing radiation detectors and measurement ranges. Sound and light alarms are activated when the preset ADER/flux density threshold levels are exceeded.
The device is included in the State Register of the Republic of Kazakhstan under no. KZ.02.01.02587-2024
Manufacturer: SOLO LLP
Country: Republic of Kazakhstan
Warranty period: 12 months
After-sales service: during the entire period of operation
The dosimeter is a wearing device designed to measure:
The dosimeter consists of a control panel and detection units depending on customer requirements. Two detection units can have separate housings or share one housing.
The flattening of nonlinearity of the energy dependence of the sensitivity of scintillators is performed by an analog-to-digital converter. The device performs basic radiation control tasks in accordance with the Hygiene Standards of the Sanitary and Epidemiological Guidelines for Radiation Safety (HS SEGRS) (NRB) and the of the Sanitary Rules of the Sanitary and Epidemiological Guidelines for Radiation Safety (SR SEGRS).
The device consists of a control panel and replaceable detection units.
Characteristic |
Value |
|
Built-in gamma X-ray radiation detection unit |
||
Detector |
scintillation type |
|
Energy range of detected gamma X-ray radiation |
from 20 keV to 10 MeV |
|
Measurement range for ambient equivalent dose rate |
from 0.01 μSv/h to 15 Sv/h |
|
Measurement range for ambient equivalent dose |
from 0.01 to 5×108 μSv |
|
Maximum acceptable reference relative error |
±12% |
|
Detection unit for beta particles and gamma X-ray radiation, external (combined)* |
||
Detector |
scintillation type |
|
Energy range of detected beta particles |
from 0.03 to 5 MeV |
|
Energy range of detected gamma X-ray radiation |
from 20 keV to 10 MeV |
|
Beta-particle flux density measuring range |
from 0.1 to 5×105 part/(min×cm2) |
|
Range of measurement of ambient dose equivalent rate of gamma X-ray radiation |
from 0.01 μSv/h to 15 Sv/h |
|
Range of measurement of ambient equivalent dose of gamma X-ray radiation |
from 0.01 to 5×108 μSv |
|
Maximum acceptable reference relative error for beta radiation |
±15% |
|
Maximum acceptable reference relative error for gamma X-ray radiation |
±12% |
|
Overall dimensions of the detector, max. |
Ø50 mm, length 140 mm |
|
Weight of detector, max. |
0.3 kg |
|
Alpha-particle and neutron radiation detection unit, external (combined) * |
||
Detector |
scintillation type |
|
Energy range of detected alpha particles |
from 1 to 10 MeV |
|
Energy range of detected neutron radiation |
from 0.025 eV to 15 MeV |
|
Alpha radiation flux density measuring range |
from 0.05 to 5×105 part/(min×cm2) |
|
Measurement range for ambient equivalent dose rate of neutron radiation |
from 0.1 to 5×105 μSv/h |
|
Measurement range for ambient equivalent dose of neutron radiation |
from 0.1 to 1×108 μSv |
|
Maximum acceptable reference relative error for alpha radiation |
±20% |
|
Maximum acceptable reference relative error for neutron radiation |
±20% |
|
Overall dimensions of the detector, max. |
Ø50 mm, length 140 mm |
|
Weight of detector, max. |
0.3 kg |
|
Gamma radiation X-ray detection unit, external (emergency) |
||
Detector |
scintillation type |
|
Energy range of detected gamma X-ray radiation |
from 20 keV to 10 MeV |
|
Measurement range for ambient equivalent dose rate |
from 0.01 to 100 Sv/h |
|
Maximum acceptable reference relative error |
12% |
|
Overall dimensions of the detector, max. |
Ø25 mm, length 130 mm |
|
Weight of detector, max. |
0.2 kg |
|
Control panel |
||
Overall dimensions of the control panel, max. |
175×125×80 mm |
|
Weight of the control panel, max. |
2.0 kg |
|
Case |
made of duralumin alloy |
|
Protection rating |
IP67 |
|
Operating temperature range of the control panel and detection units |
from –50°C to +50°C. |
|
Power supply |
built-in 7.4 V lithium-ion battery, included AC adapter with 12 V output voltage, solar cell, or car battery |
|
* It is possible to manufacture a detection unit designed to measure only one type of radiation or two separate units, depending on customer requirements |
No. |
Name |
Quantity |
1 |
Control panel with integrated gamma X-ray radiation detection unit |
1 |
2 |
Beta-particle and gamma X-ray radiation detection unit (external)* |
1* |
3 |
Alpha-particle and neutron radiation detection unit (external)* |
1* |
4 |
Emergency gamma X-ray radiation detection unit with extension cable 30 m long |
1 |
5 |
USB cable for connecting to a PC |
1 |
6 |
Power adaptera |
1 |
7 |
Extension bar, short |
1 |
8 |
Extension bar, 5 m, with case |
1 |
9 |
Headset |
1 |
10 |
Straps for moving the control panel |
2 |
11 |
Car battery charging cable |
1 |
12 |
Solar cell for charging the battery |
1 |
13 |
Case |
1 |
14 |
How-to documentation (user manual, data sheet, technical description and user guide) |
1 |
15 |
Form |
1 |
16 |
Warranty card |
1 |
17 |
Verification certificate |
1 |
* It is possible to manufacture a detection unit designed to measure only one type of radiation or two separate units, depending on customer requirements |